|
|
||||||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
General Clinical Chemistry |
1 Department of Internal Medicine; 2 Division of Nephrology and Hypertension; 3 Department of Laboratory Medicine and Pathology; 4 Division of Endocrinology, Diabetes, Metabolism and Nutrition; 5 Department of Biochemistry and Molecular Biology, Mayo Clinic College of Medicine, Rochester, MN.
aAddress correspondence to these authors at: Renal Function Laboratory, Mayo Clinic Department of Laboratory Medicine and Pathology, 200 First St. SW, Rochester, MN 55905. E-mail Lieske.John{at}mayo.edu or Larson.Timothy{at}mayo.edu.
| Abstract |
|---|
|
|
|---|
Methods: We obtained unused clinical urine samples from 150 patients who were stratified by degrees of albuminuria (<20 mg/L, 20–250 mg/L, >250 mg/L) as determined by the immunoturbidimetric assay used in our clinical laboratory (Roche Hitachi 912). Urine albumin was then remeasured via LC-MS and HPLC (AccuminTM) assays.
Results: The immunoturbidimetric assay, calibrated using manufacturer-supplied serum-derived calibrators (Diasorin), underestimated albumin compared with LC-MS. After calibration with purified HSA, this immunoturbidimetric assay correlated well with LC-MS. HPLC overestimated albumin compared with both LC-MS and immunoturbidimetry. The current LC-MS and HPLC assays both performed poorly at concentrations <20 mg/L.
Conclusions: Efforts are needed to establish gold-standard traceable calibrators for clinical assays. LC-MS is a specific method to quantify albumin in native urine when concentrations exceed 20 mg/L, and therefore could be employed for standardization among assays.
| Introduction |
|---|
|
|
|---|
Urinary albumin has traditionally been measured using antibody-based methods including RIA, ELISA, nephelometry, and immunoturbidimetry (9)(10). Measurement of urinary albumin is potentially complicated by the observation that it may exist in several forms (11). More recently, size-exclusion HPLC has been proposed as a more sensitive method to measure urinary albumin (12). In the low range (<100 mg/L), HPLC gives higher values than conventional immunoassays, and this has been hypothesized to be because of the presence of immunochemically nonreactive albumin (13)(14). Indeed, increases in urinary albumin assessed by HPLC may be an earlier predictor of a risk for progression to mature diabetic nephropathy than when microalbuminuria is measured by standard methods (12). It is also possible that size-exclusion HPLC methods may overestimate albumin concentration, since it has been demonstrated that proteins of similar molecular mass may coelute with albumin (15).
To more specifically measure urinary albumin, methods using LC-MS have been developed to precisely quantify intact albumin in native urine using N15-labeled human serum albumin (HSA) (16) and BSA(17) as internal standards. Here we evaluated an LC-MS method with an improved limit of quantification relative to the version used in the previous publications and compared it against immunoturbidimetric and size-exclusion HPLC assays at clinically relevant concentration ranges.
| Materials and Methods |
|---|
|
|
|---|
immunoassay
The routine clinical urine albumin assay in our laboratory is immunoturbidimetric (Roche Hitachi 912). We used reagents and calibrators in a kit purchased from Diasorin, Inc. (SPQ Test System). In this assay, albumin is detected using an antibody to human albumin in an automated immunoprecipitation analysis that detects turbidity and increased light scatter. Urine samples were centrifuged at 650g for 10 min before analysis. After incubation for approximately 10 min, we measured the absorbance of the solution at 340 nm. For the routine assay, we generated a calibration curve by assaying a series of manufacturer-supplied calibrators (Diasorin) with known concentrations of albumin, using the instruments data reduction capability. Concentrations for the calibrators (5.3–174 mg/L) and samples were interpolated from the calibration curve. In another set of experiments, we recalibrated the instrument using serial dilutions of HSA (Sigma-Aldrich) in charcoal-stripped urine (SeraCare Life Sciences). The stock HSA concentration (7.8 g/L) was quantified via UV absorbance (absorptivity 3.85 x 104 at 280 nm wavelength) (18).
lc-ms
We quantified urinary albumin using source-induced fragmentation as described by Loo et al. (19). In a modified version of the published protocol (17), source-induced fragmentation of intact albumin was achieved by increasing the declustering potential voltage from 100V to 336V on an Applied Biosystems API5000 triple quadrupole tandem mass spectrometer. We spiked calibrators or centrifuged urine samples (200 µL) with 20 mg/L BSA (Sigma-Aldrich), the optimal concentration to reproducibly generate an internal standard signal response without introducing interference in the HSA signal response. After the addition of BSA, all calibrators and samples were mixed vigorously for 20 min. We then loaded 60 µL sample onto an ABI POROS R2/10 (2.1 x 10 mm; 50 µm bead) column using a Thermo Scientific Aria TLX2 multiplex LC system, which allows for a vigorous cleaning cycle of the column to reduce the possibility of carryover and diversion of waste while directing sample to the MS. We performed LC for albumin separation using a 3-solvent system with a total run time of 15 min. The first 3.5 min of the LC run were sent to waste, and data were collected for the next 3.5 min, before again diverting the last 8 min of the LC run to waste. The mobile phases consisted of solvent A (water, 0.1% formic acid), solvent B (methanol, 0.1% formic acid), and solvent C (70% formic acid, 30% isopropanol). LC flow rate was 1 mL/min beginning with 95% A and 5% B for 1.5 min, ramping to 70% A and 30% B for 0.5 min, holding at 70% A and 30% B for 2 min, ramping to 50% A and 50% B over 2.5 min, then to 5% A and 95% B over 0.5 min, and holding there for 1 additional minute. The column was then cleaned at a flow rate of 2.1 mL/min for 5.5 min using 4.5% A, 0.3% B, and 95.2% C and reequilibrated with 95% A and 5% B at 1 mL/min for 1.5 min. We monitored the BSA b24+4 fragment (m/z 698.15) as an internal standard and quantitated the human albumin b24+4 fragment (m/z 685.1) using an 11-point calibration curve (quadratic fit) prepared from serial dilutions of HSA in charcoal-stripped urine as described above.
hplc
For size-exclusion HPLC (AccuminTM, Aus Am Biotechnologies), we analyzed the same urine samples with a Beckman System Gold HPLC following the Accumin protocol and using the kit components. The Accumin Direct Total Intact Albumin Assay detects total albumin after calibration with HSA liquid calibrators. Each sample was vortex-mixed for 10 s and centrifuged at approximately 2500g for 5 min. We transferred 50 µL supernatant from each centrifuged sample into separate polypropylene vials and injected 25 µL onto the column. Intact albumin was separated from other urinary proteins based on molecular size and conformation according to the package insert. Absorbance was measured at 214 nm, and we analyzed results using Beckman System Gold software.
| Results |
|---|
|
|
|---|
hitachi immunoturbidimetric assay
An operational limit of quantification was verified empirically by measuring 1 sample with a low albumin concentration (1.04 mg/L) 10 times and demonstrating an acceptable CV (9.65%). An interassay imprecision study (2 control samples analyzed over 20 days) yielded CVs of 3.1% and 1.8%, with mean values of 10.8 and 65.7 mg/L, respectively. Linearity studies, performed by diluting patient samples, yielded a linear regression equation of y = 1.0601x + 0.0609, R2 = 0.9984 over the range of 5.5 to 88 mg/L. A mean recovery of 91% was observed.
hplc
Intraassay imprecision was measured using 3 urine concentrations (19.0, 154.4, and 208.1 mg/L) assayed in triplicate on 2 different days. CVs were 0.5%, 0.6%, and 0.4%, respectively. Interassay imprecision was measured over 3 separate runs (n = 7 patient samples ranging from 18.9 to 209.3 mg/L) and resulted in CVs of 0.5%, 0.6%, and 0.4%.
method comparison
Urine albumin concentrations consistently measured higher across all concentration ranges with LC-MS compared with the Hitachi immunoturbidimetric platform using the Diasorin reagents and kit serum calibrators, as evidenced by points lying below the line of identity in Fig. 1A
. Of note, the LC-MS assay calibrators were prepared from purified commercial HSA quantitated by UV absorbance, whereas the clinical Roche immunoassay employs prediluted serum-based calibrators (Diasorin). Because the LC-MS assay cannot accurately quantify urine albumin <20 mg/L (see above), these points were deleted in Fig. 1B
before calculation of a regression line. The inset in Fig. 1B
highlights that the correlation between the 2 assays is similar in the clinically important microalbuminuric range (20–500 mg/L), as in the higher protein samples (>500 mg/L). A consistent mean negative bias of 27% in the quantifiable range for the immunoturbidimetric vs the LC-MS assay is more apparent in the Bland-Altman plot (Fig. 1C
).
|
To evaluate the potential contribution of the calibrators to the discrepancies observed between assays, we reanalyzed the same samples after the Hitachi 912 was calibrated with the purified HSA standard employed in the LC-MS assay. In this experiment, the values obtained with the HSA calibrators were higher at all concentrations compared with the Diasorin calibrators (Fig. 2A
), with the bias most apparent in the Bland-Altman plot at lower concentrations (<20 mg/L, Fig. 2B
). The breakpoint observed at approximately 200 mg/L may in part reflect the procedural detail that all samples >174 mg/L are diluted 1:10 for the Hitachi assay for quantification using Diasorin reagent B; a suboptimal curve-fitting algorithm by the instrument might also explain this result. When results for the immunoturbidimetric and LC-MS assays were compared after calibration of both assays with the same pure HSA calibrators, the correlation was improved (Fig. 3A
), without significant bias on the Bland-Altman plot throughout the range where the LC-MS assay performs well (>20 mg/L; mean 2.2% difference; Fig. 3B
). As in Fig. 2
, breakpoints are apparent, which may represent effects of sample dilution and/or the instrument curve-fitting algorithm. Therefore, discrepancies between the immunoturbidimetric and LC-MS assays appear to be largely due to calibration.
|
|
Next, we compared results from the size-exclusion HPLC assay to those from the LC-MS assay. The HPLC assay yielded higher results for albumin across all concentrations (Fig. 4A
), with the bias more apparent at concentrations <100 mg/L, and especially <50 mg/L. Overall, the mean difference with the HPLC assay was 34% (Fig. 4B
). These results are consistent with previous observations that HPLC gives higher values than immunoassay, especially in the microalbuminuric range. When the HPLC assay was calibrated using pure HSA, read-back values for the HSA calibrators correlated well with the nominal values (r2 = 0.998), although quantification was poor below 10 mg/L. Subsequently, when the Accumin serum-based standards were quantified using the HPLC assay calibrated with pure HSA, there was also a good correlation (r2 = 0.998), but values were substantially higher (approximately 30%) than the nominal values for the Accumin calibrators. Therefore, the Accumin calibrators may be another source of bias in the assay, perhaps owing to other proteins that coelute with the albumin peak in this serum-derived reagent, and contribute to discrepancies between HPLC and the LC-MS and immunoturbidimetric assays.
|
| Discussion |
|---|
|
|
|---|
Several factors could potentially contribute to the differences we observed when the immunoturbidimetric assay was performed with commercial, serum-based calibrators vs pure HSA standards (Fig. 2
). Antibodies used in the immunoassay were raised against native serum and not urinary albumin. Albumin exists in a number of isoforms, and it has been proposed that epitopes in the urine albumin molecule might not be detected by these commercially available antibodies, or that modification of albumin in the kidney might result in nonimmunoreactive albumin (12)(14)(20). However, the fact that LC-MS and immunoassay correlated quite well when the same calibrators were used (Fig. 2
) argues against this possibility. It is also conceivable that immunoassay methods could underestimate albumin concentration because of the presence of substances in the urine matrix that interfere with binding of the antigen to the antibody. If this were the case, the ratio of albumin and the interfering substance might be small at lower albumin concentrations and increase as albumin concentration rises. Therefore, one might predict better correlation between immunoassay and LC-MS at higher albumin concentrations, which we did not observe (Figs. 1
and 3
). Finally, it is possible that native urine might contain N-terminal fragments that are recognized by the LC-MS assay but not the antibody assay, accounting in part for the positive bias observed with LC-MS (Fig. 1B
). However, this bias is consistently 20% to 30% across all urinary albumin ranges, and it seems unlikely that the ratio of intact albumin to fragments would remain constant as urinary albumin increases from normal to high concentrations. Therefore, the sum of our evidence suggests that the antibodies in the Diasorin assay appropriately recognize urinary albumin. Because the current LC-MS assay does not perform well in the low to mid-normal range (<20 mg/L), our studies cannot address the performance of immunoassays in this concentration range, which likely is of clinical and epidemiological significance.
Size-exclusion HPLC has been purported to measure total urinary albumin, including immunoreactive and nonimmunoreactive albumin (13). In fact, some studies have suggested that measurement of urinary albumin by HPLC can predict the onset of overt albuminuria in type 1 and type 2 diabetics 3.9 and 2.4 years earlier, respectively, than a standard immunoassay (12)(14). It has been proposed that the epitopes on the albumin molecule might be altered by conformational changes as a result of incomplete processing by the lysosomal pathway, particularly in early diabetic nephropathy (21)(22). Other studies have suggested that size-exclusion HPLC assay can resolve albumin separately from other proteins of similar molecular size (9). In addition, a recent study confirmed that multiple proteins were present in the size-exclusion HPLC peak used to quantify albumin in urine, and additional analysis to identify proteins in the "albumin peak" using reversed-phase HPLC and MS suggested that albumin actually comprised only 70% of the material in this peak (15). In our study, HPLC gave higher results than LC-MS and immunoassay across all concentration ranges, and this effect was more pronounced at low concentration ranges (<50 mg/L). Based on the current results, we cannot exclude the possibility that nonimmunoreactive albumin exists; however, given the nature of the LC-MS assay, in order for this to be the case these immuno-nonreactive fragments would have to either totally lack (or have modified portions of) the amino-terminal 24 amino acids, or have different affinity for the reversed-phase column than the intact albumin molecule. It is still conceivable that nonimmunoreactive albumin might exist in the lowest concentration ranges (<20 mg/L), where our current study cannot compare LC-MS and HPLC.
LC-MS is a novel method for quantification of urinary albumin that relies strictly on the mass-to-charge of a specific protein fragment, whereas other traditional methods rely on antibody affinity and/or size. This LC-MS method reliably detects albumin with an intact N-terminus, and we therefore believe it provides a specific means for measurement of albumin concentration, recognizing its potential limitations. The LC-MS assay is currently not sensitive at very low albumin concentrations (<20 mg/L; Fig. 1
) owing to high signal-to-noise ratio and will need further refinements to achieve a lower limit of quantification. Because this method is relatively time consuming and requires expensive instruments, it may not be practical for routine urine albumin measurements; however, it could be invaluable to standardize between assays and calibrators that are used commonly across clinical practices. Furthermore, refinement of this assay to detect concentrations <20 mg/L in native urine, and other fragments of albumin, could allow further study into the nature, quantity, and clinical relevance of urinary albumin species in this concentration range for renal disease and cardiovascular morbidity.
| Acknowledgments |
|---|
Financial Disclosures: None declared.
Acknowledgments: The authors thank Drs. David Barnidge, Frank Crow, and Ward Lutz for helpful discussions.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
J. C. Seegmiller, D. Sviridov, T. S. Larson, T. M. Borland, G. L. Hortin, and J. C. Lieske Comparison of Urinary Albumin Quantification by Immunoturbidimetry, Competitive Immunoassay, and Protein-Cleavage Liquid Chromatography-Tandem Mass Spectrometry Clin. Chem., November 1, 2009; 55(11): 1991 - 1994. [Abstract] [Full Text] [PDF] |
||||
![]() |
J. C. Seegmiller, D. R. Barnidge, B. E. Burns, T. S. Larson, J. C. Lieske, and R. Kumar Quantification of Urinary Albumin by Using Protein Cleavage and LC-MS/MS Clin. Chem., June 1, 2009; 55(6): 1100 - 1107. [Abstract] [Full Text] [PDF] |
||||
![]() |
W. G. Miller and D. E. Bruns Laboratory issues in measuring and reporting urine albumin Nephrol. Dial. Transplant., March 1, 2009; 24(3): 717 - 718. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |