| Abstract |
|---|
|
|
|---|
Methods: We measured concentrations of hCG variants in the urine from 3 patients with apparent false-negative hCG results. Purified hCG variants were added to urines positive for hCG and tested using the OSOM, ICON® 25 hCG (Beckman Coulter), and hCG Combo SP® Brand (Cardinal Health) devices.
Results: Dilution of these 3 urine samples resulted in positive results on the OSOM device. Quantification of hCG variants in each of the 3 patient urine specimens demonstrated that hCGβcf occurred in molar excess of intact hCG. Addition of purified hCGβcf to hCG-positive urines caused false-negative hCG results using the OSOM and ICON qualitative urine hCG devices.
Conclusions: Increased concentrations of hCGβcf can cause false-negative results on the OSOM and ICON qualitative urine hCG devices. .
| Introduction |
|---|
|
|
|---|
In pregnancy, hCG immunoreactivity in urine is a result of intact hCG (intact heterodimer comprising
and β subunits) as well as partially degraded variants detectable in serum and urine. In addition to intact hCG, hyperglycosylated hCG (hCG-h),1
nicked hCG (hCGn), hCG missing the β-subunit C-terminal peptide, free β-subunit (hCGβ), hyperglycosylated free β-subunit (hCGβ-h), and nicked hCGβ (hCGβn) can be detected in both serum and urine; the core fragment of hCGβ (hCGβcf) is predominantly detected in urine (2)(3)(4)(5). The fraction of total hCG that each of these variants represents in urine changes during pregnancy, with hCG-h high in early pregnancy and hCGβcf high in midterm pregnancy (4)(5). Although most manufacturers claim that their qualitative devices are designed to detect the intact hCG molecule, little has been done to characterize the hCG variants that these devices detect. Recent reports demonstrate that many devices do recognize some of the other hCG variants (6). What is not known is whether an excess of 1 or more of these variants may cause false-negative results by binding to one of the antibodies in the devices but not the other, thus preventing formation of a complete antibody–antigen–antibody sandwich. We hypothesized that increased concentrations of 1 of the hCG variants was the cause of 3 confirmed false-negative results in a commonly used qualitative hCG device.
| Materials and Methods |
|---|
|
|
|---|
Patient 2 was a 23-year-old woman who presented to the BJH ED with sharp left flank pain. She stated that she was 10 weeks pregnant. A POC urine pregnancy test was performed in the ED (OSOM; lot 081299) and was negative. Urine was cloudy and specific gravity was 1.020, pH 7.5. There was no evidence of urinary tract infection. Ultrasound showed a live intrauterine pregnancy. Serum hCG (Siemens Centaur) was 133 860 IU/L. The patient had an MRI study to rule out appendicitis. Discharge diagnosis was low back pain. She had a medical history of UTI and subsequently came in 1 month later with threatened abortion and UTI.
Patient 3 was a 23-year-old woman who presented to the Vanderbilt University Medical Center ED with continual nausea with vomiting and abdominal pain for the past 4 weeks. She stated that she was 13 weeks pregnant. A POC urine pregnancy test (OSOM; lot 081334) performed in the Vanderbilt University Medical Center Core Laboratory was negative. Urine was orange and turbid and specific gravity was >1.035, pH 6.0. Urine was positive for trace ketones, WBCs, and bacteria. Ultrasound showed a live intrauterine pregnancy. Serum hCG (Siemens Immulite 1000, Siemens Centaur) was 99 118 IU/L. Discharge diagnosis was hyperemesis gravidarum, UTI, and hyperthyroidism.
determination of hook effect
Urine from a patient with gestational trophoblastic disease (serum hCG 3 200 000 IU/L) was serially diluted in hCG-negative urine from a male subject. We measured the hCG concentration using the Siemens Centaur total hCG assay that we have previously validated for use with urine (7). Diluted urine was tested using the OSOM device.
effect of Hcg variants on qualitative Hcg test results
We obtained purified hCG variants from the National Institute for Biological Standards and Controls (Hertfordshire, UK) [first WHO reference reagent, 2001 (hCGn, 99/642), (hCGβ, 99/650), (hCGβn, 99/692), and (hCGβcf, 99/708)]. Cross-contamination of these materials ranged from 0.004% to 1.3% (8). Each purified hCG variant was diluted in normal saline to a concentration of 3000 nmol/L, which corresponds to about 1 000 000 IU/L for intact hCG. A randomly chosen positive urine specimen with an hCG concentration of 17 800 IU/L (Siemens Centaur) was mixed with purified hCG variants at various concentrations and tested using the OSOM (lot 081299), hCG Combo (lot 8080021), and ICON 25 (lot 8029001) devices according to manufacturer instructions.
measurement of Hcg variants
We determined urine hCG variant composition using a combination of commercial and in-house immunofluorometric assays (IFMAs). We quantified intact hCG using a time-resolved IFMA specific for intact hCG and hCGn (AutoDELFIA®, Perkin-Elmer Wallac) according to the instructions of the manufacturer. This assay uses a monoclonal capture antibody specific for the β-subunit (MAb 5008) and a detection antibody (MAb 5501) specific for the
-subunit (9).
We measured the concentrations of hCGβ and hCGβcf using in-house IFMA. Both assays use the same detection antibody (MAb 1B2) that recognizes hCG, hCGβ, and hCGβcf. The capture antibody used in the hCGβ assay (MAb 9C11) specifically detects the free β-subunit of hCG (cross-reaction with intact hCG and hCGβcf is about 0.5%). The capture antibody in the hCGβcf assay (MAb 3C11) is highly specific with MAb 1B2 as tracer (cross-reaction with intact hCG and hCGβ is <0.5%). These assays were calibrated using international reference preparations (IRPs) 99/650 and 99/708, respectively. Cross-reactivities were determined as described (3) using IRPs for the various forms of hCG.
We measured total hCG (hCG + hCGβ) using the Siemens Centaur Total hCG assay. This 2-site sandwich chemiluminescent assay uses a polyclonal goat and a monoclonal mouse antibody against different epitopes of the hCG β-subunit. This assay is standardized with the WHO Third International Standard 75/537. The Bayer Centaur hCG assay does not recognize hCGβcf (10).
| Results |
|---|
|
|
|---|
hook effect on the osom device
All 3 patient urine specimens were negative when tested neat using the OSOM device but positive after dilution, initially suggesting the possibility of a hook effect (11). Parallel 10% (i.e., 90%, 80%, 70%, etc.) dilutions of urine from a patient with gestational trophoblastic disease (serum hCG 3 200 000 IU/L) demonstrated that the OSOM device produced a positive result when the hCG concentration was <1 600 000 IU/L and a negative result when the hCG was >1 900 000 IU/L. Thus, the OSOM does have a hook effect above concentrations of 1.6–1.9 million IU/L. The hCG concentration in urine from the 3 patients was far less than 1 600 000 IU/L, however, indicating that a hook effect was not the cause of the false-negative results.
results using other qualitative Hcg devices
The urine from patient 1 was tested on 2 other hCG devices. When tested on the ICON hCG device, the results were faintly positive using the neat sample and strongly positive after a 1:3 dilution into saline. When tested using the hCG Combo device, the results were clearly positive using both a neat specimen and after a 1:3 dilution. Because the hCG Combo device gives positive results in the presence of purified hCGβcf, but the ICON does not (6), this finding led us to believe that excess hCGβcf in these urines may be causing the false-negative results in the OSOM and ICON devices.
characterization of Hcg variants in patient urine
We measured hCG variants in the urine specimens from these 3 patients using variant-specific immunoassays (Table 1
). In all 3 urines, hCGβcf occurred in molar excess of intact hCG, representing 64%–78% of the immunoreactive hCG. The results for intact hCG measured by the AutoDELFIA assay were higher than those obtained by Siemens Centaur. This is likely explained by differences in calibration and/or recognition of the hCG variants occurring in urine.
|
HcgβCF inhibits test results on some qualitative devices
To further investigate the possibility that excess hCGβcf may inhibit the immunoreaction on some devices, we added purified hCGβcf to an hCG-positive urine with an hCG concentration of 17 800 IU/L as determined by the Siemens Centaur method. Results of these studies are summarized in Table 2
(and Supplemental Fig. 1, which accompanies the online version of this article at http://www.clinchem.org/content/vol55/issue7). When 1 000 000 pmol/L hCGβcf was added to hCG-positive urine, the result changed from clearly positive to negative using the OSOM device and from clearly positive to negative/faint positive using the ICON device, but no visible effect occurred on the hCG Combo device. As the concentration of hCGβcf decreased, the positive signal returned to the OSOM and ICON devices. The positive signal also returned with a 1:5 dilution in saline, as observed with the patient samples (data not shown).
|
To determine if other hCG variants might have similar effects on the OSOM device, we added purified hCGβ, hCGβn, and hCGn to hCG-positive urine (Table 3
and online Supplemental Fig. 2). At 500 000 pmol/L, hCGβ inhibited the positive signal and hCGβn and hCGn diminished the positive signal.
|
| Discussion |
|---|
|
|
|---|
hCGβcf is a fragment of hCGβ lacking the C-terminal peptide amino acids 93–145, 41–54, and 1–5 from the N-terminus (12). From approximately 5–8 weeks of gestation until term, hCGβcf is the major hCGβ-subunit–related molecule in urine and accounts for up to 90% of the immunoreactive urine hCG from midpregnancy (5). Because the concentrations of hCGβcf in serum are <0.1% of those in urine (13), it is likely that most of it is formed during excretion of hCG and hCGβ in the kidneys (14). However, it is not known why the proportion of hCGβcf in urine from some women is higher than in others.
The hook effect is a well-documented interference that occurs when extremely high concentrations of an analyte occupy all the sites on both the capture and detection antibodies and prevent the formation of a so-called sandwich in 2-site immunoassays. As a result, few or no tracer-antibody + antigen + capture-antibody complexes are formed, yielding a false-negative result. However, the measured total hCG concentrations in these 3 urines was far less than the concentrations necessary to cause a hook effect in these devices.
In these assays, we believe that hCGβcf is recognized by 1 of the 2 antibodies in the device. When the hCGβcf occurs in high concentrations and in molar excess of the forms of hCG recognized by both antibodies, it saturates one of the antibodies and blocks the ability of intact hCG to form the double antibody–antigen sandwich. This observation confirms in clinical practice the prediction of Madersbacher and Berger (15). Similar to traditional hook effects, dilution of the sample decreases the concentration of hCGβcf to a point where sandwich formation can again occur. The epitopes of various forms of hCG have been extensively characterized. The adjacent epitopes β2 or β4 on loop 1 and 3 of hCGβ occur on both hCGβcf and hCG and are used in many hCG assays (16). One of the antibodies in the OSOM and ICON assays apparently recognizes 1 of these epitopes. Other hCG variants, i.e., hCGβ, hCGβn, and hCGn, may also cause this phenomenon to varying degrees with some devices. This is probably related to the use of an antibody recognizing epitope C1 or C2. These epitopes are found only on heterodimeric intact hCG, are sensitive to nicking in hCGn, and are not present on hCGβ (16). In addition to hCG and hCG-h, the variants most frequently encountered in urine during normal pregnancy have been reported to be hCGn, hCGβ, and hCGβcf (5). A negative interference is likely to occur whenever 1 of these variants is present in great excess and recognized by only 1 of the 2 antibodies. The negative interference is more of an issue in hospital settings where qualitative devices are being used at later stages of pregnancy, when the concentrations of hCG and its variants are high.
Our results illustrate how physiological changes in the concentrations and variants of hCG in urine may cause problems for our routinely used hCG assays. The cases presented here illustrate the importance of considering the target population when designing immunoassays. Hospital and laboratory point-of-care hCG devices are not simply being used to detect early pregnancy in women trying to achieve pregnancy. They are being used in a wide variety of gestational ages. Our results show that in hospital settings, the immunoassay characteristics need to be different. Immunoassays in which hCGβcf causes a false-negative result may be acceptable for use in detection of early pregnancy but are unreliable later in pregnancy. The concentrations of hCGβcf in urine from different pregnant women vary considerably, and in addition there may be more than 10-fold differences between consecutive samples from single individuals. Compared with results from earlier studies, the concentrations of hCGβcf observed in our patients appear to represent relatively high values (17)(18). They are not unique, however; we have observed numerous urine specimens from women at 5–8 weeks gestation that exhibit faint positive results using the OSOM device but are clearly positive after dilution.
This is the first report, to our knowledge, that hCG variants occurring at high concentration in pregnancy urine cause a false-negative result in POC qualitative hCG devices. Caution should be used when hCG devices in which hCGβcf causes negative interference are used to test women who are pregnant beyond 5–8 weeks gestation, as false-negative results may occur. Devices used for this purpose should be tested to identify this problem.
| Acknowledgments |
|---|
Authors Disclosures of Potential Conflicts of Interest: Upon manuscript submission, all authors completed the Disclosures of Potential Conflict of Interest form. Potential conflicts of interest:
Employment or Leadership: None declared.
Consultant or Advisory Role: None declared.
Stock Ownership: None declared.
Honoraria: None declared.
Research Funding: None declared.
Expert Testimony: U.-H. Stenman, expert witness, Princess Diana trial.
Role of Sponsor: The funding organizations played no role in the design of study, choice of enrolled patients, review and interpretation of data, or preparation or approval of manuscript.
| Footnotes |
|---|
| References |
|---|
|
|
|---|
The following articles in journals at HighWire Press have cited this article:
![]() |
A. M. Gronowski, M. Powers, U. H. Stenman, L. Ashby, and M. G. Scott False-Negative Results from Point-of-Care Qualitative Human Chorionic Gonadotropin (hCG) Devices Caused by Excess hCG{beta} Core Fragment Vary with Device Lot Number Clin. Chem., October 1, 2009; 55(10): 1885 - 1886. [Full Text] [PDF] |
||||
| ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
| HOME | HELP | FEEDBACK | SUBSCRIPTIONS | ARCHIVE | SEARCH | TABLE OF CONTENTS |