Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 0: clinchem.2004.036954v1, 2004; 10.1373/clinchem.2004.036954
This Article
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
clinchem.2004.036954v1
50/10/1894    most recent
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Cho, O. K.
Right arrow Articles by Kuwa, K.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Cho, O. K.
Right arrow Articles by Kuwa, K.

Received on May 10, 2004
Accepted on July 26, 2004

Oak Ridge Conference

Noninvasive Measurement of Glucose by Metabolic Heat Conformation Method

Ok Kyung Cho 1*, Yoon Ok Kim 1, Hiroshi Mitsumaki 2, Katsuhiko Kuwa 3

1 PhiScience GmbH, Schwerte, Germany
2 Hitachi Ltd., Tokyo, Japan
3 University of Tsukuba, Ibaraki, Japan

* To whom correspondence should be addressed. E-mail: cho-phiscience{at}versanet.de.

Background: We developed a method, called the metabolic heat conformation (MHC), for the noninvasive measurement of blood glucose. The MHC method involves the measurement of physiologic indices related to metabolic heat generation and local oxygen supply, which correspond to the glucose concentration in the local blood supply.

Methods: We used noninvasive thermal and optical sensors on the fingertip of an individual to measure thermal generation, blood flow rate, hemoglobin (Hb) concentration, and oxyhemoglobin concentration. The calibration model incorporates mathematical procedures to convert signals from the sensor pickup to final glucose concentrations. The mathematical procedures are multivariate statistical analyses, involving values from sensor signals, polynomials from various values, regression analyses of individual patients, and cluster analyses of patient groups. The glucose value is calculated for each patient measurement, applying one of the clusters by discriminate analysis.

Results: Regression analysis was performed to compare the noninvasive method with the hexokinase method, using 127 data points (109 data points from diabetic patients, 18 data points from nondiabetic patients) with glucose concentrations ranging from ?? to ??? mmol/L (54-405 mg/dL). The correlation coefficient (r) was 0.91. Reproducibility was measured for healthy fasting persons; the CV was 6% at ??? mmol/L (100 mg/dL).

Conclusions: These data provide preliminary evidence that the MHC method can be used to estimate blood glucose concentrations noninvasively.