Clinical Chemistry Link to Randox Laboratories Web Site spacer gif
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 0: clinchem.2006.078956v1, 2007; 10.1373/clinchem.2006.078956
This Article
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
clinchem.2006.078956v1
53/5/902    most recent
Right arrow Submit an electronic Letter to
the Editor about this paper
Right arrow Alert me when this article is cited
Right arrow Alert me when eLetters are posted
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Rauh, M.
Right arrow Articles by Rascher, W.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Rauh, M.
Right arrow Articles by Rascher, W.

Received on August 24, 2006
Accepted on February 9, 2007

Endocrinology and Metabolism

Simultaneous Quantification of Ghrelin and Desacyl-Ghrelin by Liquid Chromatography-Tandem Mass Spectrometry in Plasma, Serum, and Cell Supernatants

Manfred Rauh 1*, Michael Gröschl 1, Wolfgang Rascher 1

1 Department of Pediatrics, Friedrich-Alexander-University, Erlangen, Germany

* To whom correspondence should be addressed. E-mail: manfred.rauh{at}kinder.imed.uni- erlangen.de.

Background: A sensitive method specific for ghrelins is needed for investigations of this gastrointestinal peptide. Our aim was to develop and validate a quantitative mass spectrometry (MS) method to measure ghrelin and desacyl-ghrelin simultaneously.

Methods: After deproteinization by precipitation, we performed reversed-phase separation with a rapid 2-column online extraction design coupled to a quadrupole mass spectrometer for electrospray ionization MS detection. Chromatography was performed on a C18 monolithic column, with ammonium acetate buffer/methanol as the mobile phase and a chromatographic run time of 6 min/sample. The 4-fold-charged ions were used for multiple reaction monitoring experiments.

Results: The method was linear with injections of 0.01-10 ng. Limits of detection and quantification were 0.02 and 0.07 µg/L for ghrelin, respectively, and 0.03 and 0.35 µg/L for desacyl-ghrelin. Intra- and interday imprecision (CVs) were 9%-4% and 12%-6% at concentrations of 0.33-5.93 µg/L for ghrelin, respectively, and 16%-6% and 15%-8% at concentrations of 1.12-10.02 µg/L for desacyl-ghrelin. The mean (SD) recoveries in plasma of added ghrelin and desacyl-ghrelin were 95.8% (12%) and 101% (1.2%), respectively. Using kinetic modeling, we determined the mean (SD) periods of half-change (t1/2) of ghrelin to be 156 (16) min in EDTA plasma and 49 (1) min in Li-heparin plasma. Bland-Altman analysis showed that the median differences between EIA and liquid chromatography-tandem mass spectrometry (MS/MS) for desacyl-ghrelin were -40% for plasma/serum samples and 85% for cell supernatants and for ghrelin were 6% for enriched plasma samples and 44% for cell supernatants.

Conclusion: Our HPLC-MS/MS procedure has excellent selectivity and sufficient limit of quantification to allow the monitoring of concentration-time profiles in biological matrices.