Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 0: clinchem.2007.094417v1, 2007; 10.1373/clinchem.2007.094417
This Article
Right arrow Full Text (PDF)
Right arrow All Versions of this Article:
clinchem.2007.094417v1
54/2/366    most recent
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Negi, A. S.
Right arrow Articles by Sood, A. K.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Negi, A. S.
Right arrow Articles by Sood, A. K.

Received on July 3, 2007
Accepted on November 16, 2007

Automation and Analytical Techniques

Electric Field–Enhanced Sensitivity of Grafted Ligands and Receptors

Ajay Singh Negi 1 Ajay Kumar Sood 1*

1 Department of Physics, Indian Institute of Science, Bangalore, India

* To whom correspondence should be addressed. E-mail: asood{at}physics.iisc.ernet.in.

BACKGROUND: Particle-based agglutination tests consisting of receptors grafted to colloidal microparticles are useful for detecting small quantities of corresponding ligands of interest in fluid test samples, but detection limits of such tests are limited to a certain concentration and it is most desirable to lower the detection limits and to enhance the rate of recognition of ligands.

METHODS: A mixture of receptor-coated colloidal microparticles and corresponding ligand was sandwiched between 2 indium tin oxide–coated glass plates. Electrohydrodynamic drag from an alternating-current electric field applied perpendicular to the plates increased the local concentration of the colloidal particles, improving the chances of ligand-receptor interaction and leading to the aggregation of the colloidal particles.

RESULTS: With this technique the sensitivity of the ligand-receptor recognition was increased by a factor as large as 50.

CONCLUSIONS: This method can improve the sensitivity of particle-based agglutination tests used in immunoassays and many other applications such as immunoprecipitation and chemical sniffing.