Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 0: clinchem.2008.116020v1, 2009; 10.1373/clinchem.2008.116020
This Article
Right arrow Full Text (PDF)
Right arrow Supplemental Data
Right arrow All Versions of this Article:
clinchem.2008.116020v1
55/3/573    most recent
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via HighWire
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Kavsak, P. A.
Right arrow Articles by Jaffe, A. S.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Kavsak, P. A.
Right arrow Articles by Jaffe, A. S.

Received on ,
Accepted on ,

Brief Communications

Analytic and Clinical Utility of a Next-Generation, Highly Sensitive Cardiac Troponin I Assay for Early Detection of Myocardial Injury

Peter A. Kavsak 1*, Andrew R. MacRae 2, Marie-Jeanne Yerna 3, Allan S. Jaffe 4

1 Department of Pathology and Molecular Medicine, McMaster University, Hamilton, ON, Canada
2 Department of Biochemistry and Medical Genetics, University of Manitoba, MB, Canada
3 Beckman Coulter, Inc., Chaska, MN
4 Cardiovascular Division and Division of Laboratory Medicine, Mayo Clinic, Rochester, MN

* To whom correspondence should be addressed. E-mail: kavsakp{at}mcmaster.ca.

BACKGROUND: Improvements in cardiac troponin (cTn) assays have increased the rapidity with which clinicians can identify patients with changing cTn concentrations (rise or fall) indicative of acute myocardial injury. The aim of the present study was to characterize a new, high-sensitivity cTnI (hs-cTnI) assay and examine whether increased sensitivity can result in still earlier detection of evolving injury.

METHODS: We determined the limit of detection, precision profiles, and preliminary estimates of the 99th percentile for the Beckman Coulter hs-cTnI assay in 125 healthy individuals (age <55 years, 54% male). We compared AccuTnI® and hs-cTnI to assess whether change criteria for early concentration changes (i.e., ≥3SD for low concentrations and 20% difference for concentrations ≥0.10 µg/L) were exceeded in the first 2 specimens (median time between specimens, 1 h; 25th–75th percentile, 1–3 h) from subjects with symptoms suggestive of cardiac ischemia (n = 290).

RESULTS: The limit of detection for the hs-cTnI assay was 2.06 ng/L, and the 20% CV and 10% CV concentrations were 2.95 and 8.66 ng/L, respectively. The preliminary 99th percentile estimates in lithium heparin, serum, and EDTA plasma were 9.20, 8.00, and 8.60 ng/L, respectively. In 108 patients with myocardial injury based on the peak AccuTnI concentration, applying the change criteria on the 2 earliest specimens identified 81% (95% CI 73%–88%) of patients using the hs-cTnI assay compared to 62% (53%–71%) using the AccuTnI assay (P < 0.001).

CONCLUSIONS: Although more extensive validation studies are required, this Beckman Coulter hs-cTnI assay appears to detect patients with evolving myocardial injury earlier.




The following articles in journals at HighWire Press have cited this article:


Home page
Clin. Chem.Home page
P. A. Kavsak, X. Wang, D. T. Ko, A. R. MacRae, and A. S. Jaffe
Short- and Long-Term Risk Stratification Using a Next-Generation, High-Sensitivity Research Cardiac Troponin I (hs-cTnI) Assay in an Emergency Department Chest Pain Population
Clin. Chem., October 1, 2009; 55(10): 1809 - 1815.
[Abstract] [Full Text] [PDF]


Home page
J Am Coll CardiolHome page
P. Venge, N. Johnston, B. Lindahl, and S. James
Normal plasma levels of cardiac troponin I measured by the high-sensitivity cardiac troponin I access prototype assay and the impact on the diagnosis of myocardial ischemia.
J. Am. Coll. Cardiol., September 22, 2009; 54(13): 1165 - 1172.
[Abstract] [Full Text] [PDF]


Home page
Clin. Chem.Home page
A. Pfafflin
Highly Sensitive Cardiac Troponin I Assay Leads to Lowered Specificity
Clin. Chem., September 1, 2009; 55(9): 1749 - 1749.
[Full Text] [PDF]




HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH
Copyright © 2009 by the American Association for Clinical Chemistry.