Clinical Chemistry
HOME HELP FEEDBACK SUBSCRIPTIONS ARCHIVE SEARCH
 QUICK SEARCH:   [advanced]


     


Clinical Chemistry 0: clinchem.2008.116582v1, 2009; 10.1373/clinchem.2008.116582
This Article
Right arrow Full Text (PDF)
Right arrow Supplemental Data
Right arrow All Versions of this Article:
clinchem.2008.116582v1
55/5/964    most recent
Right arrow Alert me when this article is cited
Right arrow Alert me if a correction is posted
Services
Right arrow Email this article to a friend
Right arrow Similar articles in this journal
Right arrow Similar articles in PubMed
Right arrow Alert me to new issues of the journal
Right arrow Download to citation manager
Right arrow reprints & permissions
Citing Articles
Right arrow Citing Articles via Google Scholar
Google Scholar
Right arrow Articles by Huang, Y.
Right arrow Articles by Yu, R.-Q.
Right arrow Search for Related Content
PubMed
Right arrow PubMed Citation
Right arrow Articles by Huang, Y.
Right arrow Articles by Yu, R.-Q.

Received on August 19, 2008
Accepted on January 22, 2009

Automation and Analytical Techniques

Prostate Specific Antigen Detection Using Microgapped Electrode Array Immunosensor With Enzymatic Silver Deposition

Yong Huang 1, Tai-Hong Wang 1, Jian-Hui Jiang 1*, Guo-Li Shen 1, Ru-Qin Yu 1

1 State Key Laboratory for Chemo/biosensing and Chemometrics College of Chemistry and Chemical Engineering, Hunan University, Changsha, China

* To whom correspondence should be addressed. E-mail: jianhuijiang{at}hnu.cn.

BACKGROUND: Analysis of trace proteins plays an essential role in the fields of biomedical research and clinical diagnosis. Development of methods for the detection of proteins at very low concentrations has historically been a challenge in immunochemistry. We have developed an electrical immunosensor for the detection of prostate specific antigen (PSA).

METHODS: The electrical immunosensor uses a microgapped interdigitated electrode array (MGIDEA) based on enzymatic silver deposition reaction. The deposition of silver was dispersed over the microgaps and allows the microgapped interdigitated electrodes to be electrically connected, resulting in an increase in electrical conductance of MGIDEA that is used to quantify the analyte concentration. We used this electrical immunosensor to measure PSA in human serum samples from patients with prostate diseases.

RESULTS: This electrical immunosensor exhibited a linear response with PSA concentrations over a 6-decade range from 1.0 pg/L to 1.0 µg/L, with detection limit of 0.9 pg/L. PSA concentrations using this immunosensor agreed within 10% of those obtained using a commercial chemiluminescent immunoassay.

CONCLUSIONS: The MGIDEA method has characteristics (analyte specific, low background, low limit of detection) that provide potential for molecular detection in various biomedical areas.